The “PIK-PVT” installation for reservoir fluid samples PVT studies

The PIT-PVT installation is designed to study the reservoir fluids phase behavior at conditions from reservoir to normal. The system includes a high-pressure cell with transparent window and movable piston for accurate volume and pressure measurement of reservoir mixture and test sample complete visualization. Phase boundary determination with volumes estimation is carried out using an optical-digital subsystem with a video camera. Magnetic mixer increases mixing speed and efficiency.

Research is carried out in accordance with GOST 153-39.2-048-2003.

Request product cost
Benefits/Capabilities
  • Mercury-free system
  • Automated control and data collection
  • Phase boundary visual determination
  • Oil and gas condensate research
Measured parameters
  • Gas saturation pressure
  • Oil compressibility factor
  • Oil density at reservoir pressure and temperature
  • Oil density at different stages of differential degassing
  • Volumetric oil coefficient at reservoir pressure and temperature
  • Volumetric oil coefficient at different stages of differential degassing
  • Oil viscosity at reservoir pressure and temperature
  • Oil viscosity at different stages of differential degassing
  • Gas content of oil at reservoir pressure and temperature
  • Gas content of oil at different stages of differential degassing
  • Gas condensates phase behavior studies at reservoir conditions
  • Recombined samples preparation
Technical specifications
PVT cell
Maximum working pressure, MPa 70
Maximum working temperature, °С 200
Sample volume with isolating piston, ml 130 (other volumes – on request)
Sample volume without isolating piston, ml 152 (other volumes – on request)
Viewing window size, mm 165 х 9,5
Fluid-contact parts material stainless steel
Sampling system from PVT-cell bottom and top
Cell rotation 360°
Mixer rotation frequency, rpm 300 — 2500
Fluid viscosity for mixer operation, cP at 25°C up to 100
Level control system type optical (video control)
Level control system resolution, mm (ml) 0,005 (0,015)
Temperature maintenance system
Type heating chamber, individual heaters
Operating mode heating / cooling
Operating temperature range, °C from +5 to +200
Temperature measurement accuracy, °С no more than ±1
Temperature setting resolution, °C ± 0,1
Heated elements PVT cell, supply tubes, separation tanks
Indication temperature, cell pressure
Material stainless steel