The PIT-PVT installation is designed to study the reservoir fluids phase behavior at conditions from reservoir to normal. The system includes a high-pressure cell with transparent window and movable piston for accurate volume and pressure measurement of reservoir mixture and test sample complete visualization. Phase boundary determination with volumes estimation is carried out using an optical-digital subsystem with a video camera. Magnetic mixer increases mixing speed and efficiency.
Research is carried out in accordance with GOST 153-39.2-048-2003.
PVT cell | |
Maximum working pressure, MPa | 70 |
Maximum working temperature, °С | 200 |
Sample volume with isolating piston, ml | 130 (other volumes – on request) |
Sample volume without isolating piston, ml | 152 (other volumes – on request) |
Viewing window size, mm | 165 х 9,5 |
Fluid-contact parts material | stainless steel |
Sampling system | from PVT-cell bottom and top |
Cell rotation | 360° |
Mixer rotation frequency, rpm | 300 — 2500 |
Fluid viscosity for mixer operation, cP at 25°C | up to 100 |
Level control system type | optical (video control) |
Level control system resolution, mm (ml) | 0,005 (0,015) |
Temperature maintenance system | |
Type | heating chamber, individual heaters |
Operating mode | heating / cooling |
Operating temperature range, °C | from +5 to +200 |
Temperature measurement accuracy, °С | no more than ±1 |
Temperature setting resolution, °C | ± 0,1 |
Heated elements | PVT cell, supply tubes, separation tanks |
Indication | temperature, cell pressure |
Material | stainless steel |